5-7
Operating Instructions Operator Manual 056404-443
Refer to Table 5-1 for situation when Universal Hand Control and Surgical Table Compatibility Modes do not
match.
Table 5-1. Universal Hand Control and Surgical Table — Compatibility Mismatch
Scenario* Corrective Action
1
When table is powered on, user is prompted to select [applicable] Compat-
ibility Mode.
1. Power table off by pressing
STOP button.
2. Power table on by pressing
any key except STOP.
3. Select [correct table] Mode
when prompted.
2
Control display shows BAD VERSION message but otherwise Control and
Table function correctly. When table is powered on, user is prompted to
select
[applicable] Compatibility Mode.
3 Control and Table shut down after approximately seven seconds.
1. Power table on by pressing
any key except STOP.
2. Select [correct table] Mode
when prompted.
*
Note: For scenarios 1, 2, and 3,
The Refer to Instructions symbol (at right) appears in lower right
corner of display.