4-10
056404-443 Operator Manual Operating Instructions
4.11.2 Removing the Leg Section
The surgical table leg section is designed to be easily removed (Hi-Lock
™
locking mechanism) to enable
procedures requiring optional leg supports. The leg section can be removed while a patient is on the table.
Remove the leg section as follows:
1. Using hand control, level leg section (bring to horizontal).
2. While holding leg frame, squeeze both handles and simultaneously pull backward (see Figure 4-11).
3. Safely store leg section to prevent damage.
NOTE: When leg section is removed from the table, the LCD Display shows LEG OUT if the LEG UP or LEG
DOWN hand control pushbutton is pressed.
Figure 4-11. Leg Section Removal
WARNING – PERSONAL INJURY HAZARD:
• Do not remove table leg section if patient is not properly positioned. Patient's torso and buttocks
must be firmly held in place by table back and seat sections and legs supported by leg supports.
• Do not exceed 350 lbs. (159 kg) on the leg section.