3-3
Introduction to Table Operator Manual 056404-443
3.2 Technical Specifications
3.2.1 Overall Size (W x L x H)
22 x 81 x 26 to 45"* (550 x 2055 x 675 to 1136 mm)
SAE measurements are approximate based on metric dimensions.
3.2.2 Weight
Table: Approximately 560 lbs. (250 kg)
3.2.3 Range of Motion
Height: 26 to 45" (675 to 1136 mm)*
Trendelenburg Range: 30° ±1°
Reverse Trendelenburg Range: 30° ±1°
Lateral Tilt Range: 20° ±1°
Slide Motion: HEAD: 9" (227 mm)*; FOOT: 9" (227 mm)*
Back Section Motion: UP: 80° ±1°; DOWN: 40° ±1°
Leg Motion: UP: 0° ; DOWN: 105° ±1°
Kidney Elevator Mechanism: 4" (110 mm)*
Flex/Reflex: 140° ±1° / 100° ±1°
Head Motion (Manual): 90° ±1°
3.2.4 Line Power Input: 100-240 Vac, One-Phase, 50/60 Hz, 4.0 Amp
Each table is shipped from the factory configured to the electrical requirement specified on the factory order. If
this electrical configuration needs to be changed in the field, consult STERIS for the needed procedure and
required materials.
3.2.5 Environmental Conditions
Temperature: 63-77°F (17-25°C)
Relative Humidity: 20-80% RH
Atmospheric Pressure: 700-1060 hPa
3.2.6 Transportation/Storage Conditions
Ambient Temperature: -40 - 158°F (-40 - 70°C); Not to exceed 15 weeks.
Relative Humidity: 10-100% non-condensing
Atmospheric Pressure: 500-1060 hPa