6-7
Routine Maintenance Operator Manual 056404-443
6.5 Monthly Maintenance
Check STERIS 4095 General Surgical Table casters and floor locks for any accumulated debris and clean as
follows:
NOTE: Ensure Hand Control is in proper working order and good condition.
1. Ensure floor locks are properly engaged.
2. Hold spray cleaner can 6-8" (152-203 mm) from caster and spray cleaning fluid liberally on the caster.
3. Wipe caster with a cloth, dampened with water, to remove
cleaning fluid and debris. Ensure items such as suture, oils and floor wax are removed.
4. Perform Steps 2 and 3 for remaining three casters.
6.6 Battery Charging
6.6.1 What You Should Know Before Charging the Batteries
The surgical table is equipped with INTELLIPOWER
®
dual power system. The table is powered either by facility
power or by internal batteries. If table is to be operated on battery power, note the following:
IMPORTANT: Batteries are recharged automatically as long as the Main Power Switch is set to ON and the ac
supply power is connected to the table. The PLUG icon appears on Power Panel LED Display indicating ac power
and the diode bar indicates battery condition:
When table is plugged into facility ac and Main Power Switch is ON, the diode bar shows battery charge
state. This battery charge state is also indicated by icon on primary hand control.
When table is on battery power and primary hand control is active (on), the diode bar shows battery charge
state. This battery charge state is also indicated by icon on primary hand control.
Figure 6-1. INTELLIPOWER Panel LED Display
Label 1
Label 2
ORANGE
GREEN
CHARGE
BATTERY