A-1
Appendix A Operator Manual 056404-443
EMC COMPLIANCE TECHNICAL DATA
The STERIS 4095 General Surgical Table
is designed for most surgical procedures with maximum radiological
access without patient reversing. The
STERIS 4095 General
Surgical Table and related components are intended for
use in the electromagnetic environment specified in
Table A-1
through
Table A-4
. The Customer or the user should
ensure this table is used in such an environment.
NOTE: Review all EMC safety precautions listed below before operating the STERIS 4095 General Surgical Table
within the surgical suite.
CAUTION – POSSIBLE EQUIPMENT DAMAGE:
• Appropriate components of this surgical table have been tested and found in compliance with
IEC 60601-1-2: Edition 4 2014-02, Medical Electrical Equipment - Part 1; General Requirements
for Safety; Electromagnetic Compatibility (EMC). There is, however, a potential for
electromagnetic or other interference between this equipment and other devices. Should user
experience interference, relocate this device or minimize the use of the affected equipment
while this device is in operation.
• Surgical table operation may be temporarily affected if portable or mobile RF communications
equipment is used in close proximity.
• Medical Electrical Equipment needs special precautions regarding EMC. Put table into service
according to the EMC information provided in this manual.
• Possible EMC increased EMISSIONS or decreased IMMUNITY. Do not use accessories or
replacement parts not listed in the Operator or Maintenance Manuals.
• Surgical table operation may be affected if used adjacent to or stacked with other equipment. If
adjacent or stacked use is needed, ensure table operates properly in this configuration.
APPENDIX A - EMC COMPLIANCE
A