3.2.8 Essential Performance
Essential Performance for Surgical Tables:
1) Support a patient without unwanted movement under single fault conditions.
2) Single fault conditions can include electrical, electromagnetic effects or mechanical problems.
3) In extreme cases of electromagnetic interference, operator may experience temporary loss of table function.
This situation can be corrected by removing source of interference and power cycling surgical table if necessary.
3.2.9 Image Amplification Coverage
Image amplification coverage is represented in a dimensional drawing. Refer to Figure 3-4.