A-2
056404-443 Operator Manual Appendix A
Table A-1. Guidance and Manufacturer’s Declaration – Electromagnetic Emissions – For All Equipment
and Systems (per IEC 60601-2 Clause 6.8.3.201 a – 3 Table 201)
The STERIS 4095 General Surgical Table and related components are intended for use in the electromagnetic
environment specified below. The Customer or the user should ensure this STERIS 4095 General Surgical Table is used
in such an environment.
Emissions test Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1 The STERIS 4095 General Surgical Table uses RF energy
only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference
in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The STERIS 4095 General Surgical Table is suitable for
use in all establishments, other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions IEC
61000-3-3
Complies
NOTE: The emissions characteristics of this equipment make it suitable for use in industrial areas and
hospitals (CISPR 11 Class A). If it is used in a residential environment (for which CISPR 11 Class B is
required) this equipment might not offer adequate protection to radio frequency communication services. The
user might need to take mitigation measures, such as relocating or re-orienting the equipment.