Steris 4095 - Figure 6-2. LED Relation to Battery Voltage

Steris 4095
92 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6-9
Routine Maintenance Operator Manual 056404-443
Figure 6-2. LED Relation to Battery Voltage
Recharge batteries as follows:
1. Connect ac power cord to table base and plug into an appropriate ac receptacle.
2. Turn Main Power Switch to ON.
3. Allow approximately 12 hours for full battery charge.
4. Verify FILLED BATTERY icon on hand control. Disconnect ac power.
> 24.2 V < 23.6 V < 23.0 V
< 24.2 V < 23.4 V < 22.8 V
<
24.0 V < 23.2 V < 22.6 V
<
23.8 V