5-5
Operating Instructions Operator Manual 056404-443
Figure 5-2. Universal Hand Control Screen*
*Refer to SECTION 2, HAND CONTROL SYMBOLS for definitions of symbols.
Normal Display
Error Display
ALS Display
(Auto Limit Sensor)
This normal display shows, from left to right:
Fully charged battery is powering the surgical table.
• Slide is positioned six inches toward patient’s head.
• Kidney bridge is up.
• Floor locks are engaged .
This display (error indicated by wrench symbol) shows:
• Table is powered by a low battery
• Position Sensor needs calibrated (Code 20).
• Slide is positioned six inches toward patient’s head.
• Floor locks are engaged
ALS - Display shows:
• Table operated by fully charged battery
• Suggested motion to resolve ALS condition.
• Floor Locks are engaged
Other motions can also resolve ALS condition. Generally any motion that increases separation between colliding table sections is
appropriate for resolving ALS condition.