7-2
056404-443 Operator Manual Troubleshooting
* Service charges may be incurred. Consult warranty for details.your
** Use optional foot control or auxiliary (override) control system if hand control is faulty [refer to
SECTION 5.9, BACKUP
(OVERRIDE) CONTROL SYSTEM].
No functions using hand control. Hand Control not plugged in properly – correct.
No Power – table deactivates after 30 minutes of non-
use. Press any button on hand control (except
STOP).
Faulty hand control** – replace.
Call STERIS.*
Table does not tilt right/left using hand control. Hand control not on – Press any button on hand
control (except STOP).
Faulty Hand control** – replace.
Call STERIS.*
Table does not Trendeleburg/Reverse Trendeleburg
using hand control.
Hand control not on – Press any button on hand
control (except STOP).
Faulty Hand Control** – replace.
Call STERIS.*
Table does not raise/lower using hand control. Hand control not on – Press any button on hand
control (except STOP).
Faulty hand control** – replace.
Call STERIS.*
LEVEL button on hand control does not function. All
other buttons function normally.
Sensors needed for leveling function faulty – Manually
level tabletop using other hand control function
buttons.
Faulty hand control** – replace.
Call STERIS.*
Table 7-1. Troubleshooting Chart
SITUATION ACTION