Operating Instructions Operator Manual 056404-443
18 ALARM - SLIDE sensor
failure.
Combination movements (Level, Flex and Reflex) are not
allowed. Call STERIS Service. SLIDE toward head and floor
unlock/lock commands are not allowed if SLIDE sensor is
defective. Call STERIS Service.
20 ALARM - Sensor needs
recalibration.
Combination movements (Level, Flex and Reflex) are not
allowed. Call STERIS Service.
21 ALARM - Battery fuse
missing or bad connection.
Protection fuse not installed or has failed. Call STERIS Service
to ensure fuse is installed or to replace fuse.
22 ALARM - Battery voltage
imbalance greater than 0.6.
Call STERIS Service.
23 ALARM - Charging circuit
failure (Power Supply or P6
cable or main board defect).
Call STERIS Service.
24 ALARM - Thermal protection
of Power Supply activated.
Call STERIS Service.
25 ALARM - Hydraulics failure.
No pump motor feedback
(Power Supply or P6 cable or
main board defect).
Hydraulic pump failure. Call STERIS Service.
26 ALARM - Hydraulics failure.
No valve feedback.
Hydraulic valve failure. Call STERIS Service.
27 ALARM - No slide motor
feedback.
Call STERIS Service.
28 ALARM - Continuous slide
motor feedback.
Call STERIS Service.
29 ALARM - Slide command not
received.
Call STERIS Service.
30 ALARM - Continuous slide
command received.
Call STERIS Service.
31 ALARM - Hand Control
software not compatible with
master control software.
Call STERIS Service.
32 ALARM - Solid state device
controlling power to hydraulic
pump is ON continuously.
Table turns OFF 10 seconds after message is displayed. Call
STERIS Service.
33 ALARM - Battery voltage
supplied to master control
PCB is greater than 31 V.
Table turns OFF 10 seconds after message is displayed. Call
STERIS Service.