Downloading period setting determines how often data from SV 200A will be downloaded to the PC.
Five different periods can be chosen: hourly, daily (at specified hour), weekly (at specified weekday,
at specified hour), monthly (at specified day of month, at specified hour) and custom (specified time
interval between downloads). These settings applie only when Files to download are selected.
Automatic cleaning enables choosing which files are going to be erased from the instrument’s memory
after downloading.
Cleaning period determines how often data from SV 200A will be erased. Five different periods can be
chosen: hourly, daily (at specified hour), weekly (at specified weekday, at specified hour), monthly (at
specified day of month, at specified hour) and custom (specified time interval between downloads).
This setting applies only when Automatic cleaning is active. Files are erased from the instrument only
after they had been previously downloaded with AFD.
System validation section enables configuring the following settings:
• system checks with built-in electrostatic actuator,
• validation of instrument settings after cleaning measurement files,
• validation of instrument settings after performing system check (if enabled),
• automatic synchronization of real time clock,
• verification of instrument settings.
Starting session
To start Automatic Files Download session, click Start AFD button.
The AFD session will start with global settings configured in the
Options window. Download will begin after the time defined in
Downloading period is over.