EasyManua.ls Logo

SYSMEX CyFlow Cube 8 - 2.8 Laser safety; 2.9 Electrical safety

SYSMEX CyFlow Cube 8
200 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2 Safety
CyFlow
®
Cube 8 | Operating Manual | May 2018 | Revision 032 13
2.8 Laser safety
The device is equipped with a class IIIb laser unit. If the housing of the
device is closed, the laser class of the laser unit is classified as a laser
class I (EN 60825-1:2015).
Never open the housing of the device.
Laser light of laser class IIIb can be emitted if the housing of the device is
damaged or the laser protection is removed. Serious injuries of the eyes and
the skin are the results of emission of laser light of laser class IIIb.
2.9 Electrical safety
The main electrical source used must be grounded or earthed due to local
specifications.
Please place the device in such a way that the electrical power cord can
easily be unplugged. Removable power cords should not be replaced by
inadequate wiring.
Beware of possible voltage when changing the fuse (T 4.0 A 250 V).
2.10 Alterations to the device
Unauthorised alterations to the device can result in risks and hazards.
Therefore, unless expressly permitted by the manufacturer, no alterations
may be made to the device.
If the device is altered, appropriate tests and trials must be performed to
ensure the continued safe use of the device.
2.11