6Operation
CyFlow
®
Cube 8 | Operating Manual | May 2018 | Revision 032 29
6.5 Shutdown procedure
6.5.1 Cleaning for shutdown
Perform the shutdown cleaning procedure to close the software.
Requirements
No acquisition is in progress.
Sample tube with 1.6 ml Decontamination Solution (Order No.
04-4010_R) connected.
Procedure
1. On the File tab of the Ribbon (see chapter 12.4.1 File tab on page 58)
click the Exit CyFlow Software button.
– If the Worklist or Playlist is currently open and has been modified, a
save changes dialogue appears.
– A dialogue asking for a shutdown cleaning procedure appears.
2. Select Yes in the dialogue.
– The dialogue closes and initiates the shutdown cleaning procedure.
– After the cleaning procedure is complete the software closes.
3. Remove emptied sample tube.
6.5.2 Switching off the device
Requirements
The shutdown cleaning procedure was successful and the software is
closed.
Procedure
1. In the Windows menu, click the Shut down button.
– The device shuts down.
2. Switch off the device using the power switch on the rear side of the
device.
– The device is switched off.