Presetting instrument settings
Procedure
1. Select the Settings pane (see chapter 12.5.3 Settings pane on
page 139).
2. To select a parameter as trigger parameter, set a parameter to "ON" in
the "Logic" column (only available if all other parameters are set to
"OFF").
Data of this parameter are acquired.
If desired, combine other parameters for trigger by setting them to "AND"
or "OR".
3. To predefine threshold values, enter a value in the "Threshold" column
for the relevant parameters.
4. To predefine the signal intensity to be displayed for each parameter,
enter a value in the "Voltage" column for the relevant parameters.
5. Select the Acquire pane (see chapter 12.5.4 Acquire pane on page 140).
6. Select the flow rate for the measurement by clicking the Low, Medium or
High button or by using the flow rate slider.
– The selected flow rate is displayed next to the flow rate slider.
– The instrument settings are preset.
Starting measurement of the sample
Requirements
The device is primed.
A Worklist item is active with a defined Workspace.
Procedure
1. Insert the sample tube containing the sample into the sample port of the
device until the sample tube audibly clicks.
2. On the Acquire pane (see chapter 12.5.4 Acquire pane on page 140)
click the Start button.
To refresh the view, click the Restart button on the Acquire pane.
– Data acquisition starts. Progress is displayed in the status bar.
– In the Plots tab of the Workspace (see chapter 12.8.1 Plots tab of the
Workspace on page 160) the plots display the acquired data.