8Cleaning
CyFlow
®
Cube 8 | Operating Manual | May 2018 | Revision 032 43
8.2 Performing a cleaning procedure
The device can be cleaned between sets of samples using a Cleaning
Solution (Order No 04-4009_R) followed by a sample tube with sheath fluid.
This procedure will allow you to significantly reduce cross contamination and
reduce the background. There are three different levels of intensity available
that can be set: Default, quick or deep.
Requirements
Log-in as a user with the respective role.
Procedure
1. On the Cytometer tab of the Ribbon, click the Clean button.
– An instruction dialogue is displayed.
2. Follow the instructions until the finish message is displayed.
– The device is ready for the sample measurements.