6 Operation
26 CyFlow
®
Cube 8 | Operating Manual | May 2018 | Revision 032
4. Select the respective user role.
5. Click the OK button.
– If the log-in was successful the start screen is displayed.
After a specified number of unsuccessful login attempts (default 3), the
account will be locked. If the account is locked, please contact a user with
the "user administration" privilege in its user role, e. g. a user administrator.
The default settings for logins can be changed in the user management.
Logging out
Procedure
1. On the File tab of the Ribbon (see chapter 12.4.1 File tab on page 58)
click the Log-out button.
– The current user is logged out.
– The log-in dialogue is displayed.
6.2.3 Performing a Prime
A Prime mode is available as a start-up process for the device. This process
includes cleaning, filling the tubes with sheath fluid and controlling the device
set-up.
The Prime should be performed in the following cases:
Starting the device for the first time of the day,
After the sheath fluid bottle has been filled,
As a trouble-shooting procedure (no/bad signals, blocking, etc.).
Requirements
Sample tube with 1.6 ml Decontamination Solution (Order No.
04-4010_R) connected.
Procedure
1. On the Cytometer tab of the Ribbon (see chapter 12.4.10 Cytometer tab
on page 105) or on the Acquire pane (see chapter 12.5.4 Acquire pane
on page 140) click the Prime button.
– The instruction dialogue for the Prime procedure is displayed.
2. Follow the instructions until the finish message is displayed.
– The device is ready for the sample measurements.
3. Remove emptied sample tube.