9 Maintenance
44 CyFlow
®
Cube 8 | Operating Manual | May 2018 | Revision 032
9 Maintenance
Servicing is to be performed by an authorised service engineer. Please
contact your local Sysmex representative for further instructions.
9.1
Sequence Activity Minimum
user qualifi-
cation level
Daily
maintenance
Prime the device by using the Prime
function (see chapter 6.2.3 Performing a
Prime on page 26).
Priming should be performed:
– After starting the device;
– After the sheath fluid was changed;
– After the waste bottle was emptied.
Instructed
personnel
The system should not be left electrical
switched on for several days without using
it. Shut down the system completely each
day.
Instructed
personnel
Weekly
maintenance
Clean the device according to cleaning
instructions (see chapter 8 Cleaning on
page 42).
Instructed
personnel
Quarterly
maintenance
Inline filter exchange according to
instructions.
Instructed
personnel
Preventive
maintenance
A preventive maintenance procedure needs
to be performed on a regularly basis. Please
contact the service department of your local
Sysmex representative for further
instructions.
Authorised
service
personnel