6.3 Performing a measurement
Requirements
The device is primed.
The respective Worklist is loaded.
A Worklist item is active with a defined Workspace.
The Worklist contains at least one defined item that does not contain a
listmode file.
The sample is prepared for the measurement.
The sheath fluid is not low.
The waste level is not high.
Procedure
1. Fill the sample tube with the sample fluid.
2. Insert the sample tube into the sample port of the device until the sample
tube audibly clicks.
3. On the Acquire pane (see chapter 12.5.4 Acquire pane on page 140)
click the Start button.
Data acquisition starts. Progress is displayed in the status bar.
4. If a stop condition is set, wait until acquisition stops.
If no stop condition is set, click the Stop button on the Acquire pane to
stop acquisition.
A measurement is performed.
5. Remove emptied sample tube.