6 Operation
28 CyFlow
®
Cube 8 | Operating Manual | May 2018 | Revision 032
6.4 Performing an intermediate cleaning
A cleaning mode is available to clean the device when a different sample
type is to be measured.
Requirements
Log-in as a user with the respective role.
The respective Worklist is loaded.
Sample tube with 1.6 ml Cleaning Solution (Order No. 04-4009_R)
connected.
Procedure
1. On the Cytometer tab of the Ribbon (see chapter 12.4.10 Cytometer tab
on page 105) click the Clean button.
An instruction dialogue is displayed.
2. Follow the instructions until the finish message is displayed.
An intermediate cleaning was performed.
The device is ready for the next sample measurement.