Page #241 background image
Creating and Using Math Waveforms
CSA7000B Series & TDS7000B Series Instruments User Manual
3- 173
Cursor Measurements. You can also use cursors to measure derivative wave-
forms. Use the same procedure as is found under Take cursor measurements on
page 3--181. When using that procedure, note that the amplitude measurements
on a derivative waveform will be in volts per second rather than in volt-seconds
as is indicated for the integral waveform measured in the procedure.
Figure 3- 46: Peak-peak amplitude measurement of a derivative waveform
Offset, Position, and Scale. The settings that you make for offset, scale, and
position affect the math waveform that you obtain. Note the following tips for
obtaining a good display:
H You should scale and position the source waveform so that it is contained on
screen. (Off screen waveforms may be clipped, resulting in errors in the
derivative waveform).
H You can use vertical position and vertical offset to position your source
waveform. The vertical position and vertical offset will not affect your
derivative waveform unless you position the source waveform off screen so
it is clipped.
Waveform Integration. The Math capabilities of the instrument include waveform
integration. This allows you to display an integral math waveform that is an
integrated version of the acquired waveform.

Table of Contents

Other manuals for Tektronix CSA7404B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix CSA7404B and is the answer not in the manual?

Tektronix CSA7404B Specifications

General IconGeneral
BrandTektronix
ModelCSA7404B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals