CSA7000B Series & TDS7000B Series Instruments User Manual
1-21
Incoming Inspection
This chapter contains instructions for performing the Incoming Inspection
Procedure. This procedure verifies that the instrument is operating correctly after
shipment, but does not check product specifications. This procedure contains the
following parts:
H Self Tests on page 1--22 provides instructions for performing the internal
self tests.
H Functional Tests on page 1--23 measures the time- and amplitude-reference
signals at the PROBE COMPENSATION connector.
H Perform the Extended Diagnostics on page 1--32 provides instructions for
performing internal self calibration and the extended diagnostics.
If the instrument fails any test within this section, it may need service. To contact
Tektronix for service, see Contacting Tektronix on the back of the title page.
Make sure you have put the instrument into service as detailed in Installation
starting on page 1--5. Then assemble the following test equipment and proceed
with the procedures that follow.
Assemble Equipment
Self tests do not require any test equipment. The functional tests require the
following test equipment:
H One BNC cable, such as Tektronix part number 012-0076-xx
H One1.44Mbyte,3.5inchformatteddisktocheckthefilesystem
H A P7240 probe (P7260 probe with TDS7704B )
H A probe calibration and deskew fixture, Tektronix part number 067-0405-xx
(067-0848-xx for TDS7704B)
H One TCA-BNC TekConnect adapter, or one TCA-SMA TekConnect adapter
and one SMA male-to-BNC female adapter , such as Tektronix part number
015-1018-xx