EasyManuals Logo

Tektronix CSA7404B User Manual

Tektronix CSA7404B
438 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #341 background imageLoading...
Page #341 background image
Appendix A: Specifications
CSA7000B Series & TDS7000B Series Instruments User Manual
A-7
Table A- 2: Channel input and vertical specifications (Cont.)
Characteristic Description
CSA7404B, TDS7404B, TDS7254B, &
D
S
7
1
5
4
B
2mV/divto3.98mV/div (2.5% +(6% × | net offs et/1V | ))
TDS7154B
4 mV/div to 99.5 mV/div
(2% +(2% × | net offset/1V | ))
100 mV/div to 1 V/div (2% +(2% × | net offset/10V | ))
n DC voltage measurement accuracy
CSA7404B, TDS7404B, TDS7254B, &
TDS7154B
Measurement type DC accuracy (in volt s)
Average acquisition mode (≥16 aver-
ages)
2mV/divto3.98mV/div ±[(2.5% +(6% × | net offset/1V | )) ×
| reading -- net offset | + offset accuracy
+ 0.08 division × V/division]
4 mV/div to 99.5 mV/div ±[(2% +(2% × | net offset/1V | )) × |
reading -- net offset | + offset accuracy
+ 0.08 division × V/division]
≥100 mV/div ±[(2% +(2% × | net offset/10V | )) ×
| reading -- net offset | + offset accuracy
+ 0.08 division × V/division]
Delta voltage measurement between any
two averages of ≥16 waveforms
acquired under the same setup and
b
i
t
d
i
t
i
2mV/divto3.98mV/div ±[(2.5% +(6% × | net offset/1V | )) × |
reading | + 0.16 division × V/divisi on
setting]
a
mbient
c
onditions
4 mV/div to 99.5 mV/div
±[(2% +(2% × | net offset/1V | )) ×
| reading | + 0.16 division ×
V/division setting]
100 mV/div to 1 V/div ±[(2% +(2% × | net offset/10V | )) ×
| reading | + 0.16 division ×
V/division setting]
Net offset is the nominal voltage that must be applied to the channel to bring the
trace to center screen. Net offset = offset -- ( position × volts/division) and is
expressed in volts
n DC voltage measurement accuracy,
TDS7704B
Measurement type DC accuracy (in volt s)
Average acquisition mode
(≥16 averages)
2mV/divto3.98mV/div ±[(3% +(6% × | net offset/1V | )) ×
| reading -- net offset | + offset accuracy
+ 0.08 division × V/division]
4 mV/div to 99.5 mV/div ±[(2.5% +(2% × | net offset/1V | )) ×
| reading -- net offset | + offset accuracy
+ 0.08 division × V/division]
100 mV/div to 1 V/div ±[(2.5% +(2% × | net offset/10V | ))
× | reading -- net offset | + offset
accuracy + 0.08 division × V/division]

Table of Contents

Other manuals for Tektronix CSA7404B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix CSA7404B and is the answer not in the manual?

Tektronix CSA7404B Specifications

General IconGeneral
BrandTektronix
ModelCSA7404B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals