Oscilloscope Reference Vertical Acquisition Window considerations
As you vary vertical offset, the middle voltage level moves relative to zero. This moves the vertical
acquisition window up and down on the waveform. With input signals that are smaller than the
window, it appears the waveform moves in the window. Actually, a larger signal shows what really
happens: the offset moves the middle of the v ertical acquisition window up and down on the input
signal. The next gure shows how offset moves the acquisition window to control the portion of
the wavef
orm amplitude the window captures.
Applying a negative offset m oves the vertical range down relative to the DC level of the input signal.
Likewis
e, applying a positive offset moves the vertical range up. See the following illustration.
Wh
at do you want to do next?
Learn about horizontal acquisition window considerations. (see page 684)
Learn about setting offset. (see page 487)
DSA/DPO70000D, MSO/DPO/DSA70000C, DPO7000C, and MSO/DPO5000 Series 683

Table of Contents

Other manuals for Tektronix DSA70000D Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix DSA70000D Series and is the answer not in the manual?

Tektronix DSA70000D Series Specifications

General IconGeneral
BrandTektronix
ModelDSA70000D Series
CategoryTest Equipment
LanguageEnglish

Related product manuals