RiverPro and RioPro ADCP Guide October 2017
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
Page 83
Table 10: RiverPro/RioPro Input Command Summary
Command Default Description
ZB 0 Bandwidth [0=Wide (25%), 1=Narrow (6.25%)]
ZC 064 Correlation Threshold (counts) [0..255]
ZD 111100000 Data Out {v;c;a;p;s;*;*;*;*}
ZF 020 Blanking Distance (cm) [0-500]
ZG 1 Gain [0=low, 1=high]
ZM 2 V-Beam Profile Mode [2=Linear, 12=LinPhasePlaneAvg]
ZN 010 Number of Bins [1-200]
ZO 004 Number of Mode-12 Subpings [0-20]
ZP 000 Number of Pings [0-999]
ZS 010 Bin Size (cm) [2-500]
ZV 250 Ambiguity Velocity (cm/s) [5-700]