Chapter 4 Calibration
4-9
Where:
[CO]
OUT
=
diluted CO concentration at the output manifold, ppm
[CO]
STD
=
concentration of the undiluted CO standard, ppm
F
CO
=
flow rate of CO standard corrected to 25
°
C and 760 mm Hg, LPM
F
D
=
flow rate of dilution air corrected to 25
°
C and 750 mm Hg, LPM
Allow the Model 48C to sample this CO concentration standard until a stable response is
obtained.
2. From the Main Menu select “Service Mode”. From “Service Mode” select “Lo
Multi-Point Cal”. From “Lo Multi-Point Cal” Select “Cal Point 1”. Use
and
pushbuttons to move the cursor and the
and
pushbuttons to increment or
decrement the digit values until they match the concentration being introduced to
the analyzer. Press
ENTER
. The analyzer will perform a series of calculations and
save the new parameters.
3. Press
MENU
to back up a step in the Service Mode menu.
4. Repeat step 1 for a 50% concentration of the upper range.
5. Select “Cal Point 2”
6. Use
and
pushbuttons to move the cursor and the
and
pushbuttons to
increment or decrement the digit values until they match the concentration being
introduced to the analyzer. Press
ENTER
. The analyzer will perform a series of
calculations and save the new parameters.
7. Press
MENU
to back up a step in the Service Mode menu.
8. Repeat step 1 for a 20% concentration of the upper range.
9. Select “Cal Point3”
10. Use
and
pushbuttons to move the cursor and the
and
pushbuttons to
increment or decrement the digit values until they match the concentration being
introduced to the analyzer. Press
ENTER
. The analyzer will perform a series of
calculations and save the new parameters.
11. Press
MENU
to back up a step in the Service Mode menu.
12. Select “Coefficients”.
13. Press
ENTER
. The analyzer will automatically calculate the new coefficients and
save the new parameters.
14. Repeat steps 1 through 13 for “Hi Multi-Point Cal”

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Electron 48C and is the answer not in the manual?

Thermo Electron 48C Specifications

General IconGeneral
BrandThermo Electron
Model48C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals