EasyManuals Logo

Thermo Electron 48C User Manual

Default Icon
204 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #93 background imageLoading...
Page #93 background image
Chapter 3 Operation
3-69
Temperature Compensation Coefficient
The Temperature Compensation Coefficient is used to compensate for the effects of
ambient temperatures on the zero stream. The temperature compensation coefficient can
be set by to different means, “Auto Set” which, is automatically calculated based on a
span gas reading at a previously stored temperature and the current temperature, and
“Adjust” which, is a manual adjustment made by adjusting the displayed temperature
compensation coefficient to reach the expected concentration value.
To display the Temperature Compensation Coefficient Screen:
From the Main Menu choose Service Mode
From the Service Mode choose Temp Comp Coef
Temperature Compensation Coefficient Screen
To use the Temp Comp Coef Screen:
Select either Auto Set or Adjust
If using Auto Set press ENTER. Note: The current temperature must be different
than the previous temperature.
If using Adjust use the
and
pushbuttons to match the displayed concentration
to the desired concentration.
Auto Set Screen Adjust Screen
PREV TEMP 33.1dC
CRNT TEMP 30.1dC
TEMP COEF 0.0
CALC TEMP COEF?
TEMP COMP COEF 0.0
>AUTO SET
ADJUST
CO PPM -0.108
T. COMP COEF 0.0
78 INC/DEC

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Electron 48C and is the answer not in the manual?

Thermo Electron 48C Specifications

General IconGeneral
BrandThermo Electron
Model48C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals