Chapter 3 Operation
3-25
Temperature Correction
Temperature correction provides compensation for any changes to the instrument's
output signal due to internal instrument temperature variations. The effects of
internal instrument temperature changes on the analyzer's subsystems and output
have been empirically determined. Changes in temperature are corrected by this
empirical data.
The temperature correction screen is shown below. When temperature correction
is on, the first line of the display shows the current internal instrument
temperature. When temperature correction is off, the first line of the display
shows the standard temperature of 25
o
C.
To display the Temperature Correction screen:
From the Main Menu choose Instrument Controls
From the Instrument Controls menu choose Temperature Correction
To use the Temperature Correction screen:
Use the
↑
and
↓
pushbuttons to toggle temperature correction on and off
Press
MENU
to return to the Instrument Controls menu
Press
RUN
to return to the Run screen
Temperature Correction Screen