Chapter 3 Operation
3-45
Ramp
. The digital to analog (DAC) ramp is used to fully test the analog outputs.
The analog outputs start at -2.3% and then increments by 0.1% every second until
it reaches 100.0%. A linear output indicates that the analog outputs are operating
correctly.
To display the Ramp screen:
From the Main Menu choose Diagnostics
From the Diagnostics menu choose Test Analog Outputs
From the Test Analog Outputs menu choose Ramp
To use the Ramp screen:
This is a view only screen
Press
MENU
to return to the Diagnostics menu and cancel the ramp output
Press
RUN
to return to the Run screen and cancel the ramp output
Ramp Analog Outputs Screen