Chapter 3 Operation
3-16
CALIBRATION FACTORS MENU
Calibration factors are determined during automatic and manual calibration and are used
to correct the CO concentration readings. The Calibration Factors menu displays the
calibration factors as shown below. Normally the instrument is calibrated automatically,
that is, introduce the span gas and use the Calibration Screen. However, the instrument
can be calibrated manually using this menu by changing the factors.
To display the Calibration Factors menu:
From the Main Menu choose Calibration Factors:
To use the Calibration Factors menu:
Use the
and
pushbuttons to move the cursor up and down
Press
ENTER
to go to the Calibration Factor screen
Press
MENU
to return to the Main Menu
Press
RUN
to return to the Run Screen
LO SPAN GAS 1.000
LO CO COEF 1.000
Calibration Factors Menu in Single and Dual/Autorange Modes
CALIBRATION FACTORS:
>BKG PPM 0.000
SPAN GAS 1.000
CO COEF 1.000
CALIBRATION FACTORS:
>BKG PPM 0.000
HI SPAN GAS 1.000
HI CO COEF 1.000

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Thermo Electron 48C and is the answer not in the manual?

Thermo Electron 48C Specifications

General IconGeneral
BrandThermo Electron
Model48C
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals