Chapter 7 Servicing
7-16
FUSE REPLACEMENT
Equipment Required:
115V T, 3A, 250V (Part No. 4510)
220V T, 1.25A, 250V (Part No. 14009)
1. Disconnect power to instrument.
2. Remove fuse drawer, located on the AC power connector.
3. Replace both fuses, if either is blown.
4. Insert fuse drawer and reconnect power cord.