Chapter 3 Operation
3-40
AGC Intensity
The AGC Intensity screen, shown below, displays the intensity (in Hertz) of the
reference channel Automatic Gain Control (AGC) circuit. The AGC circuit
optimizes the noise and resolution levels of the Model 48C. The AGC intensity
reading should be about 250,000 Hertz.
To display the AGC Intensity screen:
From the Main Menu choose Diagnostics
From the Diagnostics menu choose AGC Intensity
To use the AGC Intensity screen:
This is a view only screen
Press
MENU
to return to the Diagnostics menu
Press
RUN
to return to the Run screen
AGC Intensity Screen