Chapter 3 Operation
3-48
Internal Temperature
The Internal Temperature screen, shown below, displays the current internal temperature
and factory-set min and max limits. The min and max alarm limits can be changed when
the instrument is in the service mode. Acceptable alarm limits range from 8 to 47
o
C. If
the internal temperature goes beyond either the min or max limit, an alarm is activated.
The word “Alarm” appears in the Run screen and in the Main Menu.
To display the Internal Temperature screen:
From the Main Menu choose Alarm
From the Alarm menu choose Internal Temperature
To use the Internal Temperature screen:
Use the
↑
and
↓
pushbuttons to move up and down (service mode on)
Press
ENTER
to select a choice (service mode on)
Press
MENU
to return to the Alarm menu
Press
RUN
to return to the Run screen
Internal Temperature Screen