EasyManua.ls Logo

Thermo Electron 48C - CO Span Gas

Default Icon
204 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 3 Operation
3-19
CO Span Gas
The CO span gas is set prior to calibration. The span gas is used to calibrate the CO
level.
To display the CO Span Gas screen:
From the Main Menu choose Calibration Factors
From the Calibration Factors menu choose Span Gas
To use the CO Span Gas screen:
Use the
and
pushbuttons to increment/decrement the gas value
Use the
and
pushbuttons to change the gas value
Press
ENTER
to accept a change
Press
MENU
to return to the Calibration Factors menu
Press
RUN
to return to the Run screen
Single and Dual/Auto range CO Span Gas Screen
SPAN GAS:
PPM 00000.250
SET TO 00000.500
78 INC/DEC
HI SPAN GAS:
PPM 00000.250
SET TO 01000.000
78 INC/DEC

Table of Contents

Related product manuals