Chapter 3 Operation
3-61
CO Concentration
The CO Concentration screen, shown below, displays the current CO concentration and
the factory-set max alarm limit. The max alarm limit can be changed when the
instrument is in the service mode. Acceptable alarm limits range from 0 to 10,000 ppm.
If the CO concentration in single mode or low CO concentration in dual or auto range
mode exceeds the max alarm limit, an alarm is activated. The word “Alarm” appears in
the Run screen and in the Main Menu.
To display the CO Concentration screen:
From the Main Menu choose Alarm
From the Alarm menu choose CO Concentration
To use the CO Concentration screen:
Use the
and
pushbuttons to move up and down (service mode on)
Press
ENTER
to select a choice (service mode on)
Press
MENU
to return to the Alarm menu
Press
RUN
to return to the Run screen
CO Concentration Screen
CO CONCENTRATION:
ACTUAL PPM 58.3
>MAX 10000