Appendix B RS-232 Commands
B-5
set co coef
d.ddd
set low co coef
d.ddd
The “set co coef” command sets the CO coefficient to a user-defined value in the single
range and sets the high coefficient in the dual or auto range. The “set low co coefficient”
sets the low coefficient in the dual or auto range. The example below sets the CO
coefficient to 1.005.
Send: "set co coef 1.005"
Receive: "set co coef 1.005 ok"
set cal co coef
This command performs the automatic calculation of the co coefficient. The example
below shows a successful automatic calculation.
Send: "set cal co coef"
Receive: "set cal co coef ok"
set cal low co coef
This command performs the automatic calculation of the low CO coefficient. The
example below shows a successful automatic calculation.
Send: "set cal low co coef"
Receive: "set cal low co coef ok"
coef 0
coef 1
coef 2
low coef 0
low coef 1
low coef 2
The “coef 0”, “coef 1”, and “coef 2” commands report the coefficients of the curve
developed for the Hi Multi-Point Calibration and the “low coef 0”, “low coef 1”, and the
“low coef 2” commands report the coefficients of the curve developed from the Lo Multi-
Point Calibration. The example below reports the coefficient 0 having a value of 1.005
Send: "coef 0"
Receive: "coef 0 1.005"