Chapter 2 Installation
Follow the procedure below to prepare the shielded cable:
1. Remove about 1.8 cm of insulation from the cable.
2. Fold back the shielding.
3. Use electrical tape or shrink tubing to hold the shielding in place. Be sure at least 1.5
cm of shielding is exposed.
4. Strip each signal wire.
Follow the procedure below to connect the shielded cable to the 8-position header:
1. Pass the shielded cable through the cable clamp on the analog output cover, as
shown in Figure 2-4.
Figure 2-4.
Exploded View of Analog Output Cover Installation
2. Insert the bare signal wire into the slot of the header.
3. Tighten down the corresponding set screw.
4. Repeat steps 2 and 3 for each signal wire.
5. Plug the header(s) into the analog output connectors.
6. Install the analog output cover using the four #6 screws with star lockwashers.
7. Position the cable shielding so that it. comes in contact with the cable clamp.
8. Tighten down the cable clamp onto the shielding, as shown in Figure 2-5.
2-5