Chapter 3 Operation
3-9
To use the autorange mode, option switches 4 and 5 must be on. For more information
about setting the internal option switches, see “Internal Option Switches,” later in this
chapter
When the high CO range is active, the concentration must drop to 95% of the low CO
range for the low CO range to become active.
To use the autorange mode, set option switch 4 and 5 on. For more information about
setting the internal option switches, see “Internal Option Switches,” later in this chapter.
42P942-5
Figure 3-6. Analog Output in Autorange Mode