81
12 Interface
81
Interface
Chapter
12
- Read status 2
- Read status 3
* When read after digital zero is performed, it turns ON only once.
It remains OFF until digital zero is performed again.
- Write setting values
* Refer to P.83 “ ■Setting value communication format” for setting value No.
- Read setting value
* Refer to P.83 “ ■Setting value communication format” for setting value No.
- Zero calibration
Host
F325
N
RD 000 00
0
O RCRD
ID No.
ID
ID No.
Terminator
0
Hold output signal
Stable output signal
Near zero output signal
HI output signal
OK output signal
LO output signal
0: OFF 1: ON
0: OFF 1: ON
0: OFF 1: ON
0: OFF 1: ON
0: OFF 1: ON
0: OFF 1 :ON
Not defined
Host
F325
N
RE 000 00
0
O RCRE
ID No.
ID
ID No.
Terminator
0
DZ response output * 0: OFF 1: ON
Undefined
Host
F325
W CR
Setting value
Setting value
* No return data
NO
ID No.
(5-digit, no decimal point)
Sign (0, + or -)
No.
Host
F325
W
W
CR
Setting value
Setting value
NO
ID No.
ID
ID No.
Setting value
(5-digit, no decimal point)
Terminator
Sign (0, + or -)
No.
No.
Host
F325
CA
N0A
Calibrating
CRNO
ID No.
ID
ID No.
0: Normal end
2 to 7: Calibration error No.
L: Calibration LOCK (calibration not possible)
R: Calibrating
Ter min at or