User's Guide _______________________________________________________________________
98 ___________________________________________________________________ M211107EN-C
System Configuration
The SET command is used to set or update system communication and
user interface-related parameters.
The CALIBRATE command is used to set or update visibility, window
contamination, and temperature measurement calibrations. The
CALIBRATE command is available only at the advanced level. For
calibration instructions, refer to section Cleaning PWD32 on page 171.
The current system parameters can be displayed by using the
PARAMETERS command. The output is as follows:
> PARAMETERS
FS11P parameter values:
identifier: -
name:
command terminal:
lines: 24
timeout: 10
message:
type: 2
interval: 15
port: data
frame: on
simulation: off
2nd port: none
2nd type: 0
ACK/NAK: off
data port:
speed: 9600
mode: rs-232
data format:8n1
maintenance port:
speed: 9600
system:
modem: off
vis sensor: on
PWD: on
bl sensor: off
+12V out: off
PW option: enabled
The system parameter values are saved in the non-volatile FLASH
memory.
Default Factory Settings
The default factory parameter settings are listed in Table 12 on page 99.
All the listed parameters can be set to their default values with the
advanced level command SET DEFAULTS YES.