Chapter 6 _______________________________________________________________ Maintenance
VAISALA _______________________________________________________________________ 209
Replacing RAINCAP
®
1. Open the control box.
2. Unplug the rain sensor cables (X6 and X7).
3. Remove the two screws below RAINCAP
®
.
4. Remove RAINCAP
®
by pulling it out of the control box.
5. Reassemble in the reverse order. Note that the flat cable marked
with a stripe is connected to X6.
6. Calibrate the new RAINCAP
®
by giving the DRY ON command
as described in section DRY ON and WET on page 123.
0312-112
Figure 71 Removing Rain Sensor