User's Guide _______________________________________________________________________
86 ___________________________________________________________________ M211107EN-C
MESSAGE 4, Uncompensated Values
Message 4 includes both contamination compensated and uncompensated
visibility and background luminance measurement values. Message 4
also contains 3 and 10 minute average visibility values. The message
format is as follows:
FSA
VIS 02100 VUC 02100 VIS3M 02000 VIS10M 01900
AL 0 BL 01050 BUC 01050 AL 0
CSUM
First character of the unit identifier, if the ID
is not defined it is replaced by a space
character
Visibility identifier and MOR value, unit
meter (SI-unit), 1 min average
Uncompensated visibility identifier and MOR
value
Visibility identifier and MOR value, 3 min
average
Visibility identifier and MOR value, 10 min
average
Alarm status of the visibility sensor
Background luminance identifier and BL
value, unit cd/m
2
Uncompensated BL identifier and BL value
Alarm status of the background luminance
sensor
The total amount of characters in the message is 86. The transmission
times are: 2.86 s at 300 bps (10-bit char), 0.36 s at 2400 bps, and 0.090 s
at 9600 bps.
For a list of alarm status code explanations, refer to Table 11 on page 81.