User's Guide _______________________________________________________________________
120 __________________________________________________________________ M211107EN-C
Maintenance Commands
The maintenance commands are used in installation, maintenance, and
troubleshooting. They are as follows:
- STA, Displays status
- CAL Calibrator _frequency, Calibration
- CLEAN, Sets clean references
- ZERO, Displays zero and noise test status
- CHEC, Displays calibrator test signal
- DRY ON, Sets RAINCAP
®
dry offset
- WET ON, Sets RAINCAP
®
wet scale
- TCAL, Sets the TS temperature (ambient)
- HEAT, Pulls hood heater relays on
STA
The STA command (or MES 3 command) displays the results from the
built-in test system as a status message.
The system output is the following:
PWD STATUS
VAISALA PWD32 V 1.08 2010-04-26 SN:F1630103 ID STRING:
0
SIGNAL 0.49 OFFSET 147.46 DRIFT -0.55
REC. BACKSCATTER 1113 CHANGE 1
TR. BACKSCATTER -1.9 CHANGE -0.1
LEDI 4.0 AMBL -1.0
VBB 12.5 P12 11.5 M12 -11.3
TS 21.6 FS TS 23.5
TB 23
TDRD 19 21 DRD 843 839 DRY 840.9
HOOD HEATERS OFF
HARDWARE :
OK
The length of the message may vary depending on the options configured
in PWD32 and whether there are warning messages. An asterisk (*)
before a value indicates an exceeded limit.