Miscellaneous Problems 
Table 42  Miscellaneous Problems 
missing 
Check cable connections, make reset. 
parameters. 
Check the communication parameters 
(default in RS-232/RS-485: 9600/8N1) and 
the flow control parameters. 
Check the cable connections. 
To prevent collisions in half-duplex RS-485 
bus, switch line feeds off from your terminal 
If a modem is used, verify that the same 
modem standard is used in both modems 
and one is in the answer mode and the other 
Wrong polling command used. 
In the FD12 and MITRAS emulation modes, 
FS11P answers only to FD12 or MITRAS 
Open the terminal program and the 
command mode through the data or 
maintenance line and give the PARAMETER 
command. Check that the message port is 
DATA and that the data port mode is the 
Sensor is off or out of power. 
Check that the power switch is on and the 
LEDs on the FSP103 power board are lit and 
the status LED on the interface unit is 
blinking. If the AC (mains) switch is on but 
no LED is lit, check the fuses on the FSP103 
board and verify the presence of a line 
message when 
using Windows 
Select View and Font from the drop-down 
menu, and click OK. 
does not open from 
the service 
connector 
Wrong serial port on the host 
computer. 
Check the serial port your terminal software 
is using. 
Service line cabling is faulty. 
Try to open the command mode from the 
data port (connect your computer into the 
data port using the connectors on the 
FSC202 front panel) to make sure FSC202 
missing in the 
Active alarm (measurement 
value not reliable and not 
Check which alarm is active and remove the 
cause for the alarm. 
arrive but command 
mode does not open 
with the OPEN 
command. 
Use correct ID or the format OPEN * (or 
OPEN FS) that opens the command line 
regardless of the unit ID. Remove the unit ID 
parameters. 
Check the communication parameters 
(default in RS-232/RS-485: 9600/8N1) and 
the flow control parameters.