Chapter 5 _________________________________________________________ Theory of Operation
VAISALA _______________________________________________________________________ 157
Table 25 Interface Unit Event Messages
Conf0:VIS measurement only
The interface unit is configured to handle a
measurement unit sensor but no
background luminance sensor.
Conf1:ALS measurement only
The interface unit is configured to handle a
background luminance sensor but no
Conf2:VIS+ALS measurement
The interface unit is configured to handle a
measurement unit and a background
Conf8:configuration failure
The interface unit is configured to handle
neither a measurement unit nor a
background luminance sensor.
InCo1:VIS sensor not responding
Measurement unit is not responding.
InCo2:ALS sensor not
responding
Background luminance sensor is not
responding.
InCo4:PWD sensor not
responding
Present Weather Detector is not
responding.
No mains input voltage available, working
on battery.
No mains input voltage available, working
on battery.
Battery voltage is low, system will shut
CPU1:+12V output disconnected
A short circuit has been detected at the
+12 V output. The output has been shut
down.
Check connected device for short circuit.
To bring the +12 V output back into
operation, the interface unit needs a reset.
Interface unit CPU detected a memory
failure.
If the failure persists after power-resets of
the interface unit, the interface unit CPU
board has to be replaced.
internal power supply failure"
Interface unit CPU internal power failure.
If the failure persists after power-resets of
the interface unit, the interface unit CPU
board has to be replaced.
P1:PWD; cleaning required
The Present Weather Detector reports
contaminated windows.
Window cleaning is required.
P2:PWD; immediate cleaning
required
The Present Weather Detector reports dirty
windows.
Windows cleaning is required immediately.
The Present Weather Detector reports an
abnormal situation. Check status (STATUS
command) and refer to PWD User's Guide.
The Present Weather Detector is not
responding.
The Present Weather Detector is not
enabled. PW data is not available.