User's Guide _______________________________________________________________________ 
186  __________________________________________________________________ M211107EN-C 
PWD32 Calibration Procedure 
If calibrating is needed according to the calibration check, follow the 
instructions below (see section PWD32 Calibration Check Procedure on 
page 184). 
1.  Give the following command: 
 
CAL [calibrator signal value] 
 
for example: CAL  485 
 
The calibrator signal value is printed on the labels of the glass 
plates. Typically the signal is close to 500. PWD32 calculates a 
new scaling factor and stores it in the non-volatile memory 
(EEPROM). 
2.  Type CHEC to verify that the new scaling factor is in use. The 
displayed signal value should be equal to the calibrator signal 
value. If the difference between the new scaling factor and the 
factory calibrated one is more than 20%, the CAL command will be 
ignored. Check PWD32 and the calibrator for hardware or 
mechanical errors. Also consult Vaisala. 
3.  If the optical units PWC12 or PWT11 have been replaced, the new 
scaling factor might change more than 20% from the original 
scaling factor, value and the CAL command is ignored. In this case 
use the FCAL command (factory calibration) instead of CAL 
during the calibration procedure. 
PWD32 RAINCAP® Calibration 
The RAINCAP of PWD32 is calibrated as described under section DRY 
ON and WET on page 123.