User's Guide _______________________________________________________________________
180 __________________________________________________________________ M211107EN-C
Evaluating Results of Guided Calibration Procedure
for Measurement Unit
Calibration Passed
Table 31 shows an example of a successful Calibration Procedure
without re-calibration of the visibility sensor.
Table 31 Successful Guided Calibration Procedure
1> CALIBRATE FS11
Step 1/3
Clean Windows and type YES: YES
Step 2/3
Install Zero Plugs and type YES: YES
1st run
........||||||||||
Zero signal test PASSED. (0.0000)
Step 3/3
Remove zero plugs, install glass
plates and
enter signal value (between 0 and
10): 1.3497
Is signal value 1.349700 correct?
Type YES or NO? YES
1st run
..............||||
Signal test PASSED. (1.3514)
Zero Value Test and
Scatter Signal Test are passed
successfully.
Calibration procedure finished.
Calibration: PASSED.
Zero test: ok (0.0000, limit:
absolute +/-0.0001)
Signal test: ok (0.1259, limits:
recalibration +/-0.0300 rejection +/-
0.2000)
Remove calibrator assembly and type
YES: YES
1>
Guided Calibration
Procedure passed
successfully. Zero Signal
Test result was 0.0000, limit
is ±0.0001.
Scatter Signal Test
deviation is 0.1259%. Limit
for recalibration is ±0.03
(3%). Limit for rejection is
±0.2 (±20%).