There is an obstacle in the measurement
volume that causes excessive scattering.
Offset has drifted more than ±5 LSB from
zero, caused by other disturbing visibility
sensors nearby or other interference
TEMPERATURE SENSOR
FAILURE
The surface temperature sensor is
disconnected or has failed.
Table 29 Warning Messages
Window contamination has increased and
caused the window transmission to drop
below 85%. The contamination
compensation is still reliable.
Backup battery voltage low (<11 VDC),
limited operation time left
LED aged, required electrical driving power
increased, limited operation time left
Table 30 Indication Messages
AC (mains) off, hood heaters not
operational
DISCONNECTED
+12 V output on the interface unit is
disconnected due to a short-circuit in the
output line. This has no effect on the
measurement but warning lights or other
options powered by this output will not
Hood heater not operational. Hood
temperature has increased less than 2 °C
within 5 minutes since the heater was
turned on by the automatic heater control
circuit.
NOTE: If operated only with battery or DC
power, set the hood heater OFF to prevent
Dew heater not operational, no current
runs through the heater when it is on
Optical Contamination and Blocking
Monitoring
Optical contamination and blocking of both the transmitter and the
receiver are continuously monitored. The FS11P sensor monitors
contamination and blocking by measuring window contamination and
backscattered signals. The CLEAN command is used to set clean