Vaisala FS11P - Figure 49 Installing Mask Plates Onto Opaque Glass Plates

Vaisala FS11P
232 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User's Guide _______________________________________________________________________
176 __________________________________________________________________ M211107EN-C
NOTE
The tape stripes on the mask pla
tes should be placed against the glass
plates to prevent the plates from scratching the glass plates.
3. Move away from the optical path.
4. Give the CALIBRATE CHECK command. The command takes
about 90 seconds to complete.
5. If the value produced by the check command is more than 85% of
the value recorded without the mask plates, the mechanical
alignment is correct. If this is not the case, check the FS11P sensor
and the calibrator for electronic or mechanical faults. Also consult
your local Vaisala representative.
6. Disassemble the calibrator and store it into the supplied carrying
case.
7. Replace the plastic protective plate onto the calibrator assembly
seating surface on the measurement unit.
0201-078
Figure 49 Installing Mask Plates onto Opaque Glass Plates