User's Guide _______________________________________________________________________
68 ___________________________________________________________________ M211107EN-C
5. Wait for one minute and then enter the command mode with the
command OPEN. Check with the STATUS CHECK command
that no hardware errors or warnings are detected.
6. If you are connected to the data port, enter the automatic message
mode by typing CLOSE and check that a message appears on the
display every 15 seconds.
NOTE
the unit, remember to turn off both the AC (mains)
switch and the battery switch if the battery option is installed.
Initial Settings
FS11P Present Weather Sensor is typically interfaced to a host computer
or a data logger in an automatic weather observation system. After the
physical connection has been made, the details of the communication can
be configured into the FS11P software. Suitable communication settings
depend on the implementation of the entire system.
By default the sensor transmits a new ASCII data message through the
data port in RS-232 mode every 15 seconds. The user can change the
interval and message type. The sensor can also be used in a polled mode,
in which a data message is only sent when the host computer requests
one with a special command. The default communication settings are
listed in Table 8 below.
Table 8 Default Communication Settings
Polled or automatic mode, message
type
Automatic mode, message 7,
interval 15
In multipoint communication, where several sensors share the same
communication line (RS-485/Modem), the FS11P sensor should be used
in the polled mode and the individual sensors should have distinct
identifiers (ID).